mardtb  "desktop beamline"

Goniómetro plenamente motorizado con apoyo para varios tipos de detector

Chi axis

  • for easily mounting crystals without spilling liquid nitrogen
  • rotation range: 0 - 90°
  • motorized

Illumination unit

  • excellent sample recognition
  • 2 white LEDs as front light
  • 1 red LED as background light
  • independent of lab light source

Collimator

  • 2 sets of motorized slits
  • 2 ionization chambers for continuous X-ray readings
  • experimental rotary shutter at end of collimator

Cryo actuator

  • holder for various types of cold heads
  • designed for one-handed operation
  • moves cold head out of the way for crystal mounting
  • motorized version available

Crystal monitor

  • built-in camera with large LCD screen

Motorized goniometer head

  • built-in x,y,z motors
  • solenoid for holding magnetic caps
  • compatible with all types of caps
  • point & click centering by software

Distance & 2-theta

  • translation along beam approx 380 mm
  • 2-theta rotation of up to 30°
  • mount for many types of detectors (up to 80 kg)

Automatic alignment system

  • hor. & vert. motor for translations around beam
  • hor. & vert. motor for rotations around beam
  • fully automatic beam search
  • fully automatic beam optimization
  • continuous monitoring of X-ray beam

Con el goniómetro mardtb traemos todas aquellas características agradables que los usuarios conocen de las líneas de luz de sincrotrón al laboratorio domestico. Ofrece un grado muy alto de automatización. Prácticamente todos los movimientos son motorizados. Todos los motores son controlados por un microprocesador que puede ser operado a través de una interfaz Ethernet desde un ordenador.

Opcionalmente, el mardtb puede ser controlado desde un dispositivo móvil (Android, iOS próximamente). Consulte la página de la aplicación mardtbapp para obtener más información.

HIGHLIGHTS

VERSÁTIL

El instrumento se puede combinar con cualquier fuente de rayos X desde los tubos sellados Mo hasta los últimos sincrotrones. El instrumento aloja el detector de placa de imagen mar345, el detector CCD SX-165 / MX-225 o la linea de detectors de Dectris PILATUS4 (260k/1M) y EIGER2 (1M/4M)

ALINEACIÓN AUTOMÁTICA DEL HAZ

Gracias a un soporte cardánico, a las cámaras de ionización integradas y a la motorización total, el instrumento se alineará automáticamente con el haz pulsando un solo botón en el software suministrado.

GEOMETRÍA PHI INVERTIDA

El motor de rotación CHI incorporado permite que el eje PHI que sostiene la cabeza del goniómetro se desplace hacia arriba en una posición vertical. Esto hace que el cryo-montaje sea realmente sencillo.

VÍDEO INTEGRADO

Un área de aprox. 2 x 1,5 mm alrededor de la posición de la muestra se mostrará en el monitor LCD incorporado sentado en la parte superior del colimador. El flujo de vídeo se redirige a un equipo de control.

VERDADERO EJE 2-THETA

Un motor 2-theta permite una rotación vertical de 30 grados de un detector alrededor de la muestra.

SIN HIERVA

El instrumento proporciona un soporte fijo para una amplia gama de dispositivos de crio-refrigeración (Oxford Cryosystems, Oxford Instruments, Rigaku Xtreme, etc.). Un escape de cryo-flujo con caudal ajustable asegura un flujo laminar de la corriente fría alrededor de la muestra y elimina la formación de hielo.

HASTA 22 MOTORES

Dependiendo de la configuración del mardtb (vea más abajo), se pueden encontrar hasta 22 motores en el instrumento.

DEPENDABILIDAD

Tras muchos años de uso intensivo en muchos sincrotrones y laboratorios domésticos, podemos afirmar que no existe nada comparable a la fiabilidad del goniómetro mardtb.

COSTE DE PROPIEDAD

El goniómetro mardtb es famoso por su fiabilidad. El coste de propiedad del mardtb, en particular, se aproxima a cero.

CONFIGURACIONES

mardtb desktop beamline with mar345 image plate detector

14 motores en configuración estándar

El mardtb comprende motores para un eje PHI (rotación de muestra), un eje Z (traslado de muestra horizontal), un CHI-swing, una traslación del detector con respeto a la muestra, una etapa 2-teta, una viga y dos pares de rendijas horizontales y verticales. El sistema de alineación cardánica agrega 4 motores más para traslados y rotaciones del instrumento.

mardtb desktop beamline with mautomatic goniometer head

Cabeza de goniómetro automático: 16 motores

The automatic goniometer head adds x- and y-translation translations to the goniometer head (the z-translation being a standard component), making a standard IUCr-type goniometer head obsolete. Sample centering can be done from the computer or via a local motor control box.

mardtb easymount: automatic sample loading mechanism

easymount: 19 motores

El easymount es un mecanismo de montaje y recuperación de muestras completamente motorizado. El easymount agrega un motor para el traslado del cryo-refrigerador y 2 motores para el sistema de transporte de la muestra a la configuración automática de la cabeza del goniómetro.

marcsc automatic sample changer

Cambiador de muestras: 22 motores

El cambiador de muestras añade 3 motores más a la configuración de easymount para el almacenamiento de muestras.

OPCIONES & ACCESORIOS

mardtb desktop beamline with automatic goniometer head Cabeza automática del goniómetro con traslados x, y, z motorizados:
  • Motores paso a paso para las traslados x, y, z- de la muestra montada
  • Solenoide para sujetar tapones magnéticos
  • Software con rutinas de búsqueda y centrado de muestras sin intervención humana, así como rutinas de centrado de point y click
Motorized cryo slider Cryo-slider motorizado: 
  • Slider para mover automáticamente la cabeza del enfriador cryo 20 mm hacia arriba o hacia abajo a lo largo de la corriente fría
Illumination unit for better sample alignment Unidad de iluminación de muestra: 
  • LLED para crear condiciones de iluminación homogéneas para el sistema integrado de visualización de muestra
  • Paleta rotativa para crear un fondo homogéneo
mardtb app  mardtb app: 

ESPECIFICACIONES

Características generales
  • Stand alone operation
    • Works with any X-ray source (Mo sealed tube to latest synchrotrons)
    • Detector mounts for: mar345, Rayonix SX-165 | MX-225, Dectris EIGER2 | PILATUS3
    • Compatible with mar345s, Rayonix and Dectris detector systems
  • Automatic alignment via host computer
  • Dedicated microprocessor
    • Real-time OS, TCP/IP services
    • 10BaseT Ethernet interface
    • Industrial serial bus for expandability
    • Open software interface for easy customization
Sistema colimador
  • Motorized slits
    • Automatic beam shape optimization
    • Variable aperture from 0 to 5 mm with 2.5 µm resolution
    • Individual settings for vertical and horizontal beam width
    • Tungsten slits for high quality beam shaping
    • Automatic or manual setting
  • Ionization chambers with large dynamic range (108)
    • Covers sources from 3rd generation synchrotrons to sealed tubes
    • Fast automatic beam optimization and analysis
    • Automatic capture of primary beam characteristics
  • Fast rotary beam shutter
    • Rotary solenoid for fast operation (1 ms)
    • Optical open- and close-state verification
    • Adaptive exposure synchronization
  • High resolution CCD microscope
    • Viewing direction along the beam path (20° offset)
    • Highly sensitive high resolution CCD camera (0.0003 lux / f1.4)
    • 2 mm x 2 mm viewing area for easy crystal location and centering
    • 4" LCD monitor integrated in collimator (70:1 magnification)
    • Video output for secondary monitor and frame grabber
    • Crystal view capture during data collection
  • Motorized precision tungsten primary beamstop
    • Automatic beamstop to crystal distance optimization
    • Easy beamstop alignment aided by microscope
    • Electronic detector collision protection
    • Easily attachable/detachable
    • 50 mm maximum crystal to beamstop for low resolution optimization
  • Fast rotary beam shutter
    • Rotary solenoid for fast operation (1 ms)
    • Optical open- and close-state verification
    • Adaptive exposure synchronization
Sistema goniométrico
  • High precision phi-axis
    • Horizontal configuration
    • Resolution: 0.000625°/step, speed: up to 900°/min
    • Accuracy: 0.002°, reproducibility: 0.00125°
    • Manual gear unlock (anticlockwise) for easy crystal centering
    • Electronic referencing of zero position
    • Exposure synchronization with shutter feedback
    • Enhanced IUCR goniometer head mount
  • Motor driven phi-axis translation
    • Automatic/manual adjustment, electronic position referencing
    • 20 mm range for a 49 mm IUCR goniometer head, resolution: 0.3125 µm/step
    • Positioning speed up to 12.5 mm/sec for cryogenic purposes
  • Motorized phi-axis swing ('inverted phi')
    • Integrated into phi-axis assembly
    • Configurable swing angle from 0° to 90°
Montaje del detector
  • Motorized vertical 2-theta arm (0° to 30°)
    • mar345: 2-thetamax = 96°, dmin(MoKalpha)=0.48 Å, dmin(CuKalpha)=1.04 Å
    • SX-165: 2-thetamax = 106°, dmin(MoKalpha)=0.44 Å, dmin(CuKalpha)=0.96 Å
    • EIGER2 R 4M: 2-thetamax = 98.2°, dmin(GaKalpha)=1.20 Å, dmin(CuKalpha)=1.35 Å
    • No distance restrictions, electronic position referencing
    • High resolution positioning (approx. 0.0001°/step at about 1°/sec)
  • Motorized crystal to detector distance translation stage
    • mar345: 75 mm to 425 mm, marCCD: 20 mm to 370 mm, Dectris EIGER2 R 4M: 35 to 390 mm
    • Integrated into 2-theta arm , no theta restrictions
    • Electronic position referencing
    • Resolution: 1.25 µm/step, speed: up to 25 mm/sec
Sistema de alineación
  • Positional alignment
    • High precision vertical and horizontal translation
    • Automatic primary beam profiling
    • Vertical range: ± 7.5 mm (352.5 mm to 367.5 mm above table top)
    • Vertical resolution: < 1 µm (0.05 µm/step), vertical speed: 0.2 mm/sec
    • Horizontal range: ± 10 mm
    • Horizontal resolution: < 1 µm (0.3125 µm/step), horizontal speed: 3 mm/sec
  • Positional alignment
    • Directional alignment
    • Spindle driven horizontal and vertical rotation
    • Automatic beam optimization
    • Vertical rotation range: approx. ± 2°
    • Vertical resolution: approx. 0.3 µrad/step, vertical speed: approx. 2 mrad/sec
    • Horizontal rotation range: ± 3.5°
    • Horizontal resolution: approx. 0.8 µrad/step, horizontal speed: approx. 15 mrad/sec
  • Beam Profiling
    • Configurable adaptive beam optimization:
    • any beam from any optic
    • Fully automatic 3D beam profile analysis
Accesorios criogénicos
  • Cryostream mount at 45° about the beam axis (5° towards the source)
  • Cryostream exhaust with an adjustable flow rate, up to 50 l/minute capacity and virtually no icing near the sample
  • Support for Oxford Cryostream 600 to 800 & Cobra series, Oxford Instruments Cryojet and Rigaku Xstream
Requisitos informáticos
CPU Pentium Pro 200Mhz or better (Intel, AMD, etc.)
Memory (RAM) >= 512 MB
Disk space >= 20 GB
Video card: Support for 24-bit colors at >= 1280x1024 pixels
Network card: Dedicated >= 10 Mbit/s with IP-address 192.0.2.2
TV-card: Optional feature for mardtb goniometer (see below)
Supported OS: Linux: Ubuntu
On request: Mac OS X, Windows, other Linux flavours
Required OS components X-windows, X11-environment, openmotif libraries

 

  • Un PC DEBE estar equipado con una tarjeta Ethernet dedicada configurada en la dirección 192.0.2.2.
  • Se recomienda encarecidamente un capturador de fotogramas de vídeo. Una memoria USB de bajo coste es suficiente, pero asegúrate de que el sistema operativo la soporta.